.. _user-manual-metal-transfer: Move the probe from the case to the stereotax ============================================= For surgery, the metal holder moves from the case onto the stereotactic adapter, with drive and probe still mounted. The starting point is a probe stored in the closed case (end of step 6 in :ref:`user-manual-metal-load-store`). .. important:: Do this next to the stereotactic frame at your surgery setup, so the adapter goes straight from the case into the stereotax. .. admonition:: You will need - The closed case with the loaded probe - Stereotactic adapter with connector basket - Blue rod support block - 1.5 mm hex driver Step 1: Open the case --------------------- .. figure:: ../../../../_static/images/r2metal/r2metal_T1_open_case.png :width: 100% :alt: Transfer step 1: loosen the two captive panel screws and lift the lid off. #. Loosen the two captive panel screws. #. Lift the lid off. Step 2: Lay the stereotactic adapter in the case ------------------------------------------------ .. figure:: ../../../../_static/images/r2metal/r2metal_T2_adapter_into_case.png :width: 100% :alt: Transfer step 2: place the blue block, then lay the 8 mm rod into the case. #. Place the blue rod support block at the end of the case. #. Lay the 8 mm rod in so that it rests in the groove of the blue block. #. The lower end of the rod sits in the adapter channel of the case, next to the holder. Step 3: Move the holder to the stereotactic adapter --------------------------------------------------- .. figure:: ../../../../_static/images/r2metal/r2metal_T3_holder_to_adapter.png :width: 100% :alt: Transfer step 3: move the holder with drive and probe onto the adapter. #. Take the connector/EIB out of the case basket. #. Unscrew the metal holder from the case and move it, with drive and probe, onto the stereotactic adapter. #. Fix it with the same M3 panel screw. #. Fix the connector/EIB in the adapter's connector basket. Step 4: Tension the flex cable on the adapter --------------------------------------------- .. figure:: ../../../../_static/images/r2metal/r2metal_T4_tension_flex_on_adapter.png :width: 100% :alt: Transfer step 4: loosen the sleeve set screw, slide the basket up, tighten again. #. Loosen the set screw on the blue sleeve. #. Slide the connector basket up until the flex cable is taut. #. Tighten the set screw again. .. note:: The sleeve set screw does the same job as the upper set screw in the case. Step 5: Ready for the stereotax ------------------------------- .. figure:: ../../../../_static/images/r2metal/r2metal_T5_ready_for_stereotax.png :width: 100% :alt: Transfer step 5: the adapter carries holder, drive and probe. The stereotactic adapter now carries the holder, drive and probe, with the connector/EIB in its own basket. #. Take the adapter out of the case. #. Clamp the 8 mm rod in the stereotax. In the stereotactic frame ------------------------- The 8 mm rod clamps where the frame's standard electrode holder would sit. Because the clamp screw is tightened from above, the holder stays clear of ear bars, bite bar and neighbouring implants. .. figure:: ../../../../_static/images/r2metal/r2metal_A6_stereotax.png :width: 100% :alt: Both holders in a rodent stereotactic frame, with close-ups of the implant site.